An artifact is any signal on the EEG that does not come from the brain. Spotting artifacts, fixing the ones you can and noting the rest is one of the most important parts of an EEG tech's job.
Last updated . Written by EEG Diagnostic Tech.
Why artifacts matter
Some artifacts look like brain activity. If they are not fixed or noted, they can be mistaken for abnormal patterns. The tech is the only person who sees the patient during the recording, so the tech's notes tell the physician what was happening at that moment.
Patient artifacts
- Eye blinks and eye movements: large slow waves at the front of the head, in the frontopolar channels. Note them, and ask the patient to keep the eyes closed and still when needed.
- Muscle: fast, spiky activity, often from tense jaw or forehead muscles. Ask the patient to relax the jaw. The high-frequency filter (often 70 Hz for routine EEG) reduces very fast activity, but turning it down too far can hide real activity.
- Sweat: slow sway in the baseline. Cooling the patient and the room helps; the low-frequency filter (often 1 Hz) reduces very slow activity.
- Photomyogenic response: frontal muscle twitches time-locked to the photic flash. It is artifact, not brain activity.
Electrode and equipment artifacts
- Electrode pop: a sudden sharp deflection from a loose or poorly prepped electrode. It shows in every channel that shares the bad electrode. Check its impedance and re-prep or re-apply it.
- 60 Hz (power-line) noise: a thick, fast band, also in the channels that share the bad electrode. Check impedance first. The notch filter removes 60 Hz noise in North America; use it only when needed, and note it.
- Salt bridge: paste smeared between neighboring electrodes shorts them, so their channel looks nearly flat. Clean the paste and re-apply.
How to find the bad electrode
In a bipolar montage, look for the channels that share one electrode: a pop or 60 Hz noise appears in both. Check impedance before you log an artifact or re-prep. A common impedance target is below 5 kΩ, with the electrodes balanced.
EEG artifacts quiz and practice
Recognizing artifacts takes practice on real-looking traces. In our online EEG technician program, Chapter 2 covers identifying artifacts, the EEG Tech Room explains every artifact as you record, and members can race practice opponents in Artifact Sprint in the Competition Room. See the EEG simulator and practice games (members).
Questions
What is an EEG artifact?
Any signal on the EEG that does not come from the brain, for example eye blinks, muscle activity, a loose electrode or power-line noise.
How do you fix electrode pop?
Find the electrode shared by the affected channels, check its impedance, and re-prep or re-apply it.
Should I use the 60 Hz notch filter?
Check impedance and fix the electrode first. Use the notch filter only when needed, and note that it was used.
What causes a flat channel?
A common cause is a salt bridge: paste smeared between neighboring electrodes, which shorts them. Clean the paste and re-apply the electrodes.
More answers on the FAQ page.